The Statim S-cycle Cassette Autoclave sterilises handpieces in only 8 minutes.
Think how much time and money that'll save you.
|
The Statim 2000 S-cycle cassette autoclave is the fastest benchtop steriliser available. Statim will enable you to sterilise your handpieces in only 8 minutes and between patients. This will optimise your clinical time management and improve efficiency.
It will also dramatically reduce the number of handpieces you will require in the average day from around seven to three; saving you money.
The gentle sterilisation of the Statim and the reduced time for which your valuable handpieces are exposed to the sterilising process, will also prolong their useful life and reduce maintenance costs.
It’s a new formula! Speed equals money saved.
|
How Does Statim Work So Fast?
|
|
The removal of air has long been recognized as critical to achieving sterilisation because air can act as a barrier to the sterilising steam.
In fact, unless the autoclave can evacuate air from inside the handpiece before sterilization begins; the handpiece would not necessarily be sterilised.
This ‘pre-evacuation’ and sterilisation can only be assured in a validated S or B-cycle autoclave.
|
The Certainty Of Validation
|
The Statim S-cycle cassette autoclave is a bespoke steriliser in that it is validated microbiologically by independent research institutes and tested for specific instruments. These include all major brands of handpieces.
Now you can also obtain daily validation of Statim’s sterilisation cycle with a specifically designed Process Challenge Device – StatimHelix that makes Statim an S class autoclave that can prove its effectiveness by both biological and mechanical means.
And with daily validation, you can be certain that Statim is meeting the challenge of ensuring that yourself, your staff and your patients are protected against cross-infection from contaminated instruments.
Statim also validates its own successful cycle. A computerized microprocessor aborts the cycle if any of the parameters of temperature, pressure or time are not achieved. You can be certain that by following the simple keypad instructions the fully automatic operation will result in sterilisation of all your solid, hollow, wrapped or unwrapped instruments.
|
Easy To Use And Maintain
|
Very simple keypad operation and helpful LED display prompts makes Statim S-cycle cassette autoclaves easy to use and user friendly.
Statim is also self-diagnosing – if a problem occurs it will immediately be indicated on the electronic display panel… very reassuring.
|
 |
 |
 |
Start |
Stop |
Level Indicator |
| One touch begins the sterilization process. |
Manual override allows safe interruptio at any stage of the process. |
A cleverly designed spirit level indicator assures set-up and completely safe operation. |
|
The Statim Difference
Attractive, compact, reliable and gentle
|
The Statim S-cycle cassette autoclave is a compact autoclave small enough to fit neatly into a convenient space on, or under, the dental cabinetry countertop as it generates no heat and very little noise. It is an attractive, well-designed, aesthetic unit which sits comfortably with today’s surgery décor.
Statim prolongs the life of expensive and delicate instruments such as fibre-optic dental handpieces, orthodontic pliers, endodontic instruments, syringe tips and diamond and carbide burs. Statim’s steam injector system provides temperature consistency and expels the air that causes oxidation in conventional autoclaves. Statim cassette autoclaves use only clean distilled water with every cycle, preventing the formation of rust and the build-up of bio contaminants. Also, a built-in air pump flushes the chamber with filtered dry air, further reducing humidity.
Light, thin-walled Statim cassettes promote rapid heating and cooling, further minimising overall processing time and reducing instrument exposure to heat. All designed to maximise the useful life of your valuable instruments and to minimize maintenance costs.
|
|
 |
 |
 |
| Ultra-thin walled cassettes allow for rapid heating and cooling. |
Precision inlet/outlet system for quick, thorough steam injection and air purging. |
Schematic of forced air removal design with combined purging. |
|
 |
 |
 |
| Very simple operation. Just select the cycle and press start. That’s all there is to it! |
Accurate record keeping and tracking is provided by the Statim 5000 integral printer or an optional external printer for Statim 2000 |
An important biological filter will not allow bacteria or spores above 0.3 µm to enter the sterilising chamber during the drying cycle. An efficient air filtration system includes easy access filters for quick replacement |
| Specifications table for: |
Statim 2000 |
Statim 5000 |
| Machine Dimensions: |
Length: |
48.5 cm |
55 cm |
| Width: |
41.5 cm |
41.5 cm |
| Height: |
15 cm |
19 cm |
Cassette Size (External): (Includes Handles) |
Length: |
41 cm |
49.5 cm |
| Width: |
19.5 cm |
19.5 cm |
| Height: |
4 cm |
8 cm |
| Cassette Size (Internal): |
Length: |
28 cm |
38 cm |
| Width: |
18 cm |
18 cm |
| Height: |
4 cm |
8 cm |
| Sterlisation Chamber Volume: |
1.8 l |
5.1 l |
| Reservoir Volume: |
4.0 l |
4.0 l |
| Weight (Withough Water): |
21 Kg |
33 Kg |
| Clearance required: |
Top: |
50 mm |
50 mm |
| Sides: |
50 mm |
50 mm |
| Back: |
50 mm |
50 mm |
| Front: |
48 cm |
57 cm |
| Clearance required to remove cassette: |
48 cm |
57 cm |
| Minimum charge in the water resevoir: |
550 ml |
550 ml |
| PRV value (pressure relief valve): |
3 BAR |
3 BAR |
| Electrical Rating: |
220 - 240V, 50Hz, 6A |
220 - 240V, 50Hz, 6A |
| Current: |
AC |
AC |
| Protection Class: |
I |
I |
| Protection: |
Covered |
Covered |
| Ambient Operating Temperature: |
5°C - 40°C |
5°C - 40°C |
| Sound Levels: |
Mean - 56 dB Peak - 65 dB |
Mean - 56 dB Peak - 65 dB |
| Humidity: |
80% Max. |
80% Max. |
| Max. Altitude: |
2000 m |
2000 m |
| Optional printer Statim 2000/5000 |
| Type: |
Thermal Printer |
| Print: |
20 characters per line |
| Print Speed: |
1 line per second |
Paper Roll Capacity: approx. |
80 sterilisation cycles per roll |